2/27
C 50
Alcatel Vacuum Technology France - ASM 180 TD/TD+ - ASM 181 TD+ User’s Manual
Alphanumeric Control and Display
Panel (ACDP) operation
Edition 03 - May 97
Automatic operation
Purpose of the ACDP
option
Manual operation
This option is intended for
industrial inspection.
It is used to:
• display messages concerning the status of the detector;
• display the measurement in digital form;
• configure and control the unit in two operating modes
(manual or automatic);
• print inspection tickets using an external printer connected to
the associated interface (see
B 40
).
• To monitor the audio signal according to the selected
configuration
• The operator retains control of the test cycle start and stop
using the cycle key
or the
key on the ACDP.
• The Red and Green indicator lights (rejected part / accepted
part) come on according to the value of the Helium signal in
relation to the programmed manual reject level.
• The choice of test mode (gross leak, fine leak or sniffer) is up
to the user.
• The detector can be used as if it did not include this option
(see standard detector operation in
C 20
).
• The audio signal is working on a FIXED or FLOATING mode,
according to the selected configuration.
• The test cycle is automated according to the programmable
test parameters.
• It is possible to program up to 10 references of different
parts each with its own test parameters.
• At the end of each automatic cycle, the test result is
displayed, the part is sorted as "Accepted" or "rejected"
according to the programmed reject level.
• An audio signal could be activated, according to the result of
the test and the selected configuration.
C
The ACDP panel operating instructions are given in the following pages.
We advise you to read them carefully in order to become familiar with its
operation.
At the end of C 50 (pages 26 and 27), the user will find a view of the front panel
of the option and a flow chart showing all the available functions.