47
The special cable is
disconnected.
Connect the special cable correctly.
For details, see
Connecting to your personal
computer.
The data read software is not
installed on the personal
computer.
Install the data read software.
For details, see the instruction manual for software.
The communication cannot
be operated.
The communication settings are
made incorrectly.
Check the input port.
See the instruction manual for software.
The display goes out.
The auto power off is activated.
Make the auto power off inactive.
For details, see
Various settings-3, Setting the
auto power off function.
The display is dark.
The LCD contrast is low.
Adjust the LCD contrast.
For details, see
Various settings-5, Adjusting the
LCD contrast.
The message, “Need to
change a backup battery !”,
is shown.
The backup battery runs out.
Contact your local sales representative or sales
office.
Check the above items and take corrective actions. If any malfunction still occurs, stop operation of