34
8. Default settings of recording
Log mode
Tag mode
The default settings are set. The default settings are convenient for the settings of recording which is
different from the initial settings (factory default settings) and may be used frequently, for example the
input types.
This setting is made in the same manner as described for
section 5, Setting for the recording.
8-1 Display the home screen and press
Clear
FUNC
key.
8-2 Select [Other settings] and press
ENT
key.
8-3 Select [Default settings] and press
ENT
key.
(The screen is changed to the setting screen of sensor type)
Log mode
Tag mode
In this unit, the “K”, “T”, “E”, or “J” type thermocouple are available.
Additionally, when using the optional voltage input adaptor or current input adaptor (sold separately),
it is available for DC voltage (0 to 5V) or DC current (0 to 20 mA) signals.
Move the cursor to a channel to set by the [
] or [
] key, and change the input type by the [<] or [>]
key.
Symbol
K
T
E
J
V
−
Input type “K” thermocouple
“T” thermocouple “E” thermocouple
“J” thermocouple
Voltage/Current
adaptor
No
connection
⇒
Press [CANCEL] key to return to the previous screen.
CAUTION
•
When the input signal type is set to “V”, the unit and scale can be set.
•
If
“
−
“ (no connection) is set for all channels, the data cannot be recorded.
ログ 番号
各 チャン ネルの 入力種 類
Log No.
Sensor type of each channel