46
Troubleshooting
Trouble Cause
Corrective
action
The batteries are not put
correctly.
Put the batteries correctly.
For details, see
Mounting and replacing the
batteries.
The batteries run out.
Replace the batteries with new ones.
For details, see
Mounting and replacing the
batteries.
The screen is not displayed.
The AC adaptor is disconnected.
Connect the AC adaptor correctly.
For details, see
Using the AC adaptor.
The sensor is disconnected.
Insert the sensor connector correctly.
For details, see
Connecting the sensor.
The current/voltage input lines
are disconnected.
Connect the input lines correctly.
For details, see
Connecting the sensor.
The unit is set to “
−
“ (no
connection).
Change the setting of input type.
For details, see 5. Settings of the recording.
(Operation manual).
The over-range trouble occurs.
Check the excessive input or sensor type.
For details, see 5. Settings of the recording.
(Operation manual).
The display of recorded
value is “
−
“. No numeric
values are shown.
The sensor is faulty.
Replace the sensor with a new one.
The thermocouple type is
different.
Set the thermocouple type correctly.
For details, see 5. Settings of the recording.
(Operation manual).
The offset value is changed.
Change the offset value.
For details, see various settings-4. Setting offset
values.
The temperature is
measured in error.
The sensor is faulty.
Replace the sensor with a new one.
The strong static electricity,
electromagnetic wave, or
high-frequency may adversely
affect the unit.
Do not put this unit or sensor close to an equipment
producing the strong static electricity or
electromagnetic wave.
The indication value
fluctuates largely.
The sensor is faulty.
Replace the sensor with a new one.
All settings are set at “
−
“ (no
connection).
Set the input type correctly.
For details, see 5. Settings of the recording.
(Operation manual).
The data cannot be
recorded.
The remaining memory capacity
is insufficient.
Save the recorded data and delete the recorded
data.
The recording mode is changed.
The recorded data does not
exist.
The recording setting is
changed.
Save the recorded data certainly before changing
the recording mode or recording setting.
Name of working person or
remark cannot be displayed.
The Name of working person or
remark is not registered or
selected.
Register or select the working person or remark.
For details, see 6. Recording personnel and
comment(Operation Manual).