20 

 

Optical Fusion Splicer Instruction Manual 

 

Fig 5.2.1 Install heating sleeve 

Fig 5.3.1 Strip coating length 

 

 

5.3 Strip and Clean the Optical Fiber Coating 
a. 

Strip the coating 

Strip the fiber’s coating with a stripper, the length is about 
30mm, as Fig.5.3.1&5.3.2 shows. Place the coating stripped fiber 
into the fixture groove, the length is about 30mm. 

 

 
 

 
 
 
 
 
 

Summary of Contents for SAT-17T

Page 1: ...ual SAT 17T OPTICAL FUSION SPLICER Before operating the equipment please carefully read this instruction manual Do follow all safety instructions and warnings covered in this manual Take the manual fo...

Page 2: ...nt is subject to change without notice All rights reserved Reproduction adaptation or translation of this material is prohibited without prior written permission of Guanggu Inc except as allowed under...

Page 3: ...ND INTERFACE INSTRUCTIONS 13 1 Power Supply 13 2 Power On Off 15 3 Main Menu 15 4 System Setting and Function Setting Menu 17 5 Preparation Before Splicing 18 6 Splicing Operation 24 7 Tension Test 31...

Page 4: ...F SAT 17T Fusion Splicer Instruction Manual APPENDIX 1 WARNING INFO LIST 42 APPENDIX 2 ERROR MESSAGE LIST 47 APPENDIX 3 GUARANTY 49...

Page 5: ...sideration to user s personal safety and injury and provided a lot of Safety Cautions instructions Misuse of this machine may result in electric shock fire and or serious personal injury Read and unde...

Page 6: ...t damage also may result in fire personal injury or death Please use the original supplied AC power cord and do not place heavy objects on the AC power cord Fuming electric shock or equipment damage m...

Page 7: ...ectrical current may cause personal injury burning electric shock and other damages for equipment Do not touch the splicer AC power cord and AC plugs with wet hands These may result electric shock Do...

Page 8: ...y area where temperature and humidity are extremely high otherwise it may result equipment failure Do not touch sleeve directly or heater during heating or immediately after completion of heating Thei...

Page 9: ...tion Replace the electrodes must in a pair Ignore or misunderstanding to follow the above instructions may cause unusual arc discharge It can result equipment damage or degradation in splicing perform...

Page 10: ...unique functions especially suitable for optical fiber cable works and maintenance in telecommunications broadcasting railways petrochemical power military public security and other communication fie...

Page 11: ...oped fiber Average Loss 0 02dB SM 0 01dB MM 0 04dB DS 0 04dB NZDS 0 02dB BIF UBIF Return Loss Better than 60dB Typical Splicing Time 9s 7s fast mode Fiber Aligning Core alignment and cladding alignmen...

Page 12: ...ustable Power supply Built in lithium battery 11 1V AC DC adapter input AC 100 240V Output DC 13 5V 5 0A Working environment Temperature 20 C 55 C 4 F 131 F Humidity 95 RH 40 C non condensing Altitude...

Page 13: ...Chargeable Battery 1 pc 3 AC DC Adaptor 1 pc 4 AC Power Cable 1 pc 5 Tweezers 1 pc 6 Cooling Tray 1 pc 7 Carrying case 1 pc 8 Fiber Strippers 1 pc 9 Indoor Fiber Stripper 1 pc 10 Alcohol Bottle 1 pc...

Page 14: ...al Fusion Splicer Instruction Manual 4 Description and Function of Splicer 4 1 Components of Splicer Heating tube LCD Screen Power Port USB Port Wind Protector Battery Fig 4 1 1 Main host Fig 4 1 2 Po...

Page 15: ...Splicer Instruction Manual Fig 4 1 4 Heating stove Fixture Wind protector Electrodes V groove LED Light Electrodes cover Fig 4 1 3 Splicing components Heat V groove Heat V groove Optical fiber press f...

Page 16: ...Function Start stop Power on off Heating key Start heating Switch X Y view field Switch the X Y views after fiber alignment Menu Confirm Enter menu and confirm the operation Reset Motor Reset Start S...

Page 17: ...Charge This fusion splicer adaptor input index wide range 100 240V 1A 50 60Hz Output index 13 5V 5A please must use the dedicated adaptor Connect fusion splicer and power source with adaptor if batter...

Page 18: ...too low to run normally a warning will show as Fig 1 1 4 user must charge for the battery or connect adaptor to supply power d Battery usage precaution 1 Before first time use please fully charge the...

Page 19: ...ptical fiber checking interface as fig 2 2 1 shows The power model will be auto judged while the battery is used to supply power the surplus electricity will be Showed 2 2 Power Off Press in long time...

Page 20: ...sting etc Splice Mode Setting parameters under splicing mode Heating Mode Setting parameters of heating process time temperature heating shrink etc History Records Record total ARC splice ARC reset ti...

Page 21: ...guage such as Chinese English etc Screen Reversal Screen can rotate in 180 to suit different direction working Set Time Set the system s time year month date etc Factory Restore Recover all the system...

Page 22: ...on splicer auto to reset waiting time Turn on tension test this can t run Auto Start If close cover set turn on after closing the cover auto align and splice Auto Heating If close cover set turn on af...

Page 23: ...ner plastic with stripper 4 Stripped coating with stripper 4 Strip coating with stripper Fig 5 1 2 Strip other type of optical fiber 5 2 Install Heating Sleeve After the optical fiber splicing a heati...

Page 24: ...eve Fig 5 3 1 Strip coating length 5 3 Strip and Clean the Optical Fiber Coating a Strip the coating Strip the fiber s coating with a stripper the length is about 30mm as Fig 5 3 1 5 3 2 shows Place t...

Page 25: ...ter stripping the coating use gauze dipped in alcohol cotton ball to clean the bare fiber toward to bare fiber and rotate the ball to clean the chips As fig 5 3 3 5 4 Cleave the Optical Fiber a Ensure...

Page 26: ...olding optical fiber by another hand flick off fragmentary optical fiber and carefully remove the optical fiber from cleaver e Take out and place the fiber chips into chips collector box f Indoor fibe...

Page 27: ...ts so that optical fibers end surface quality can be guaranteed c Hold fiber with fingers and close fiber clamp tip cover to let optical fiber in a fixed position of v groove Make sure fibers are plac...

Page 28: ...ice Program Edit the current item s splice parameters as Fig6 1 1 Cleaning ARC Time Clear ARC is cleaning the dust on optical fiber surface by ARC The clean time range 0 200ms End face Angle Limit Set...

Page 29: ...ick Splice Mode Can be set On off if open splice time will at min 7 sec Forced Splice Mode When aligning the fiber if the fiber can t calibrate normally such as cleave end face isn t meet the align ne...

Page 30: ...ioanl ARC time ARC2 Current Set additioanl ARC current 6 2 Auto align and End face Check To ensure excellent splicing this fusion splicer is adopted Image processing System to check the optical fiber...

Page 31: ...auto reset motors user can also press to reset motor and re cleave or re load fiber Image X Y axis Prompt Info Probable Reason Measureme nt Incorrectly placed right fiber Right fiber is not load into...

Page 32: ...face Problem occur while cleaving fiber concave core bulge tip bevel burr etc Re cleave fiber Unqualified fiber Dust on the surface of fiber Re clean fiber and reload fiber 6 3 Splicing After fiber a...

Page 33: ...ce of splice loss Splice result normal but the splice loss exceed limitation after splicing user should press to re ARC after ARC system will re check the fiber evaluate splice loss and show whether i...

Page 34: ...ase splice current or time Fiber Separation 1 Low fiber push value 2 Low speed of fiber push 3 Higher splice current or longer splice time 1 Execute Detect system parameter 2 Decrease Prefuse current...

Page 35: ...truction Manual Fig 7 1 Tension Test Menu Fig 8 1 History Records Menu 7 Tension Test If Tension Test is turn on after finishing splicing Tension Test will be executed automatically Tension 2N the men...

Page 36: ...ount the ARC total times from last Reset Reset ARC Counter Change the electrodes execute this operation Total Spliced The statistic of stored splice records Splice Records Review Provide max 10000 gro...

Page 37: ...Mode as Fig 9 1 shows Heating Mode Description Heating Mode No Different heating mode is set for different sleeve 30 group user define program Sleeve Type 10mm 60mm sleeve FC SC Sleeve Diameter 1 8mm...

Page 38: ...o be straight and keep the sleeve in the center of heating stove then close the heater cover d Choose heating mode and confirm heating parameters If same with previous heating ignore this step e Press...

Page 39: ...ARC current Clean Electrodes Run repeatedly high current ARC to clean electrodes Replace Electrodes After replace electrodes auto detect ARC location and ARC repeatedly to stabilize electrodes Detect...

Page 40: ...to splice spot axis skewing before after ARC to judge the ARC strength and auto calibrate it to standard strength thus to realize a fiber splice in low loss and high stabilization Strongly suggest Ru...

Page 41: ...ibration mode 3 Electrodes Maintenance 3 1 Clean Electrodes ARC may be effected by the impurity on the electrodes surface so user have to clean electrodes periodically Operation step as below a Press...

Page 42: ...the new electrodes and its cover tighten the screw d Check whether two electrodes are on a horizontal plane if not install the electrodes once again e Power on and load the fiber into splicer in Maint...

Page 43: ...pt info on screen if failed accord to prompt to re start system detect Step b e After finish successfully press to exit Detect System Parameters mode 5 Clean Fusion Splicer 5 1 Clean V groove In most...

Page 44: ...a dry cotton to wipe out the residual alcohol check the objective lens glass and confirm it is cleaning totally d Power on and check the screen to confirm whether there is still dust if yes repeat the...

Page 45: ...ith the current version if yes then user can ignore this operation c Power on the fusion splicer use USB wire to connect the fusion splicer and PC after the computer pop up the U disk prompt copy the...

Page 46: ...n Detect System Parameter or contact with factory Right fiber placed incorrectly RFPC 1 Right fiber is cleaved too short 2 Right fiber loaded in V groove is break 3 Right fiber isn t loaded in V groov...

Page 47: ...value Re cleave left fiber if the quality still poor replace cleaver Note Menu Splice Mode End face Angle Limit to set the upper limit value Right fiber end face unqualified RFEANQ Right fiber end fac...

Page 48: ...s quantity is exceed limited value Download the original splice records by USB then contact with Authorized company to operate Reset Counter Abnormal aligning 1 Dust on end face or bad quality end fac...

Page 49: ...talled the wind protector cover Execute Detect System Parameter if the problem isn t solved contact with factory Abnormal detecting Malfunction during run Detect System Parameter Check whether the fib...

Page 50: ...ed or the relevant connector loosen Re start the fusion splicer if the prompt still show contact with factory directly Abnormal sensor The inner senor works abnormally The abnormalities do not affect...

Page 51: ...ctrodes tip or replace electrodes When do ARC fusion splicer crash Incorrectly install the electrodes Re install electrodes strictly comply with request ARC Calibration failed Current environment affe...

Page 52: ...ng 2 Screen ribbon cable loosen or damage Power off then re start fusion splicer if the problem is not solved contact with factory directly The fibers aren t spliced after running splicing ARC strengt...

Page 53: ...ation manual or operation instruction Wearing part such as electrode etc Note In order to ensure service quality before sending back the splicer user is required to contact sales agency or manufactory...

Reviews: