12 Technical description
Choice of protocol
From the “Admin Config” menu, select the protocol which corresponds to the external device, using the
encoder wheel:
•
OTP,
•
Monnal Link,
•
VueLink/Intellibridge EC10.
Contact the information system manager to select the appropriate protocol.
Connecting the communication module
Connect the cable supplied by the manufacturer of the communication module to RS232 connector
no. 1 on the
Monnal T75
and then connect the other end of the cable to the external device.
Do not leave the cable connected to the
Monnal T75
without connecting it to an external device, in or-
der to prevent electromagnetic interference.
Air Liquide Medical Systems protocols
The
Monnal T75
includes the following Air Liquide Medical Systems communication protocols:
•
OTP,
•
Monnal Link.
These protocols are available upon request.
For more information, contact [email protected].
Information transmitted
The information transmitted is as follows:
•
ventilation set-points,
•
alarms,
•
alarm thresholds,
•
all measurements apart from time curves.
Compatible communication modules
•
Capsuletech Datacaptor (OTP protocol),
•
Philips monitors (VueLink/Intellibridge Protocol EC10),
•
Mindray (OTP protocol).
To obtain a full list of the communication modules compatible with the
Monnal T75
, contact the manu-
facturer.
A hospital network failure can result in no data being transmitted (alarms and monitoring) to the exter-
nal device.
Connection to a hospital network which includes other apparatus could give rise to previously unidenti-
fied risks to patients, operators or third parties.
The user should identify, analyze, evaluate and control such risks. Subsequent modifications made to
the hospital network could introduce new risks and require additional analysis. Modifications to the
hospital network include:
•
modifications to the configuration of the external device;
•
the connection of additional units to the external device;
•
the disconnection of units from the external device;
•
the updating of devices connected to the external device;