12 Technical description
12.1 Operation
12.1.1 Pneumatic system
AIR
Ambient air intake
PATIENT
Patient
O2 LP
Low-pressure oxygen inlet (concentrator)
NEBULISEUR
Nebulization outlet (calibrated coupler) to
which the nebulizer is connected
O2 HP
High-pressure O2 inlet
(network, cylinder)
BF1
Bronze filter (laminar flow)
PS1
O2 pressure sensor
C1
Non-return valve
PS2
Differential pressure sensor
C2
Non-return valve
PS3
Balloon pressure sensor
C3
Non-return valve
F1
Air intake filter (microfilter)
PS4
Airway inspiratory pressure sensor
F2
O2 inlet filter (oxygen concentrator)
PS5
Airway inspiratory pressure sensor (redun-
dancy
F3
O2 inlet filter (compressed gas)
T
Turbine
F4
Air intake filter
TS1
Patient gas temperature sensor
FS1
O2 flow sensor
V1
Oxygen regulating proportional solenoid
valve
FS2
Blower flow sensor
V2
Nebulization on/off solenoid valve
FS3
Patient flow sensor
V3
PEEP control proportional solenoid valve
FS4
Hot wire expiratory flow sensor
V4
PEEP control on/off solenoid valve
102