Monnal T75 user manual
N°
Alarm
Priority
Alarm activation
Alarm onset delay
Actions
98
Nebulization due to start
INFO
Nebulization due to
start
Triggering 15 min
before the start of a
nebulization cycle
99
Screen locked
INFO
The user has locked
the screen
Immediate triggering
100
Unit on
Stand-by
INFO
Unit in stand-by
Immediate
triggering at ventila-
tion stop or at the
starting of the device
101
FiO2 sensor calibration: plan to
restart the automatic tests!
LOW
[acknowledgeable]
Deviation from the
O2 sensor measure-
ment
Immediate triggering
102
Suction; patient disconnection in
progress
MEDIUM
Disconnecting the
patient with the “Suc-
tion” function active
Triggering within 15
s
Validate to acknowl-
edge the alarm
Conduct the auto-
matic test as soon
as possible
103
Oxygenation in progress – Suction
possible
INFO
Press the “xx%
O2↗”
button
Immediate triggering
Reconnect the pa-
tient
104
End-of-suction oxygenation in
progress
INFO
Reconnecting the cir-
cuit after aspiration
Triggering within 10
s
105
Expi. flow measurement disabled,
use an external monitor
INFO
Disable the hot wire
sensor
Immediate triggering
Use an external mon-
itor
106
Low minute volume alarm disabled
LOW
Immediate triggering
Set a low MVe alarm
threshold different
from OFF
107
Press detected on \nthe touch
screen!
LOW
Electronic fault
Immediate triggering
Change the ventila-
tion and contact the
tech. department
Exhaled volume < Vt
target and Delivered
pressure = Ppeak
Immediate triggering
Check that settings
of Vt target and
Ppeak are consistent
108
Vt target and set pressure incompat-
ibility!!
MEDIUM
Exhaled volume > Vt
target and delivered
pressure = PEEP + 5
cmH2O
Immediate triggering
Check that settings
of Vt targe (Pressure
delivered cannot be
lower than PEEP
+ 5 cmH2O)
109
Oxygen therapy enabled, Apnea venti-
lation disabled
INFO
Press the
'Oxygen
therapy'
button
Starts after pressing
the
'Start'
button
110
Oxygen therapy branch may be ob-
structed!!!
HIGH
Pressure in pa-
tient circuit > alarm
threshold
Starts after 4 s
Check the patient cir-
cuit, the patient/ma-
chine interface and
the inspiratory limb
of the ventilator.
Check that the alarm
levels match the set-
tings.
111
R and C stat measurement enabled
INFO
Press the 'R&C stat'
button
Immediate triggering
112
'P0.1 measurement enabled'
INFO
Press the 'P0.1' but-
ton
Immediate triggering