4 Use
•
Do not connect the IRMA™ at the outlet of an elbow connector, in order to avoid an accumulation
of patient secretions in the adapter and the obstruction of XTP
™
windows.
•
When using the IRMA™ probe, place the adapter so that its XTP
™
windows are vertical, to avoid an
accumulation of patient secretions on these windows:
•
When using the IRMA™ probe, position the adapter so that moisture and secretions do not accu-
mulate inside it due to the effect of gravity. Replace the adapter if it has been rained on or if it is
damp.
•
Do not use any aerosol medication with the IRMA™ probe, to avoid a deterioration of IR transmis-
sion through the windows of the adapter.
•
If the calibration is not done properly, the measurement values will be skewed.
•
Never sterilize the IRMA™ probe or immerse it in liquid.
•
Do not pull on the cable of the IRMA™ probe.
•
Do not use the IRMA™ probe at a temperature below 0°C or above 40°C.
•
During use, ensure that the IRMA™ probe is connected to the ventilator before you connect it to the
patient.
•
The probe can be cleaned using a cloth dipped in alcohol (maximum 70% ethanol or 70% iso-
propanol).
•
Remove the adapter before cleaning the probe.
•
The adapters are not autoclavable.
•
The gas measurements supplied by the probe must be checked regularly using a reference instru-
ment. We recommend that this check be performed annually.
4.10 Display of pressure and flow rate curves
The pressure and flow rate curves are displayed in real time by default. The volume and CO2 curves
may also be displayed (if the software option is present).