Monnal T75 user manual
3.
Press the
[Back]
key to quit the change curve menu.
Adjustment of scales
To adjust the time scale, press the
[s]
key.
To adjust the pressure scale, press the ordinate axis of the pressure curve or the
[cmH2O]
key.
To adjust the flow rate scale, press the ordinate axis of the flow rate curve or the
[L/min]
key.
To adjust the volume scale, press the y-axis of the volume curve or the
[mL]
key.
To adjust the CO2 scale, press the y-axis of the CO2 curve, the
[mmHg]
key, the
[kPa]
key or the
[%]
key.
4.9.4 Setting up the CO2 measurement probe
1.
Connect the IRMA™ probe to the etCO2 connection socket
(1)
.
1
2.
Apply power to the ventilator.
3.
Connect the probe to its patient adapter (a). The probe is correctly connected to its adapter
when you hear it click into position.
4.
The LED flashes and then goes green. This means that the IRMA™ probe is ready for use (b).
5.
Connect the IRMA™ probe, equipped with its adapter, to the Y-piece on the patient circuit (c).
6.
Connect the IRMA™ probe to the patient’s endotracheal tube (d).
7.
Position the IRMA™ probe (e) (see photo opposite).
Note:
The probe sends information and alarms to
Monnal T75
. The probe has an LED indicat-
ing the following statuses: