11 Questions and Answers
- 92
–
11.2 How to reset the Data Logger Settings
If you follow the following steps, all settings made so far will be
deleted.
After the reset, the saved measurements will no longer contain start
and stop times (i.e. no timestamps will be available).
With the reset taking effect, the time and date in the data logger will be
set to 00:00 (time) and 00.00.2000 (date).
Before resetting the data logger, export the saved measurements to
keep the start and stop times.
The pairing data of the connected wireless sensors will not be deleted.
!
The data logger must be switched off.
1. Press and hold the Hotkey
6
, while pressing the ON/OFF key
7
. Press
and hold those keys until the home screen will be displayed.
➢
All settings made so far will be deleted.
➢
The data logger is switched on.
2. Set the time and date of the data logger before starting a new
measurement (see chapter 8.1.7 How to set Time, Date and Time
Zone).