7 Usage
- 39
–
featuring the following information: Alarm occurrence time, time of
alarm end, and the corresponding measuring channel. In addition, an
acoustic signal will sound (provided that the signal transmitter has
been switched on) and the ALARM LED
5
will light up.
Recognize an alarm
If an alarm is triggered in the data logger, a dialog field informs you
about the new entry of an alarm in the alarm list. The alarm dialog
can be switched off (see chapter 8.1.16 How to switch the Alarm
Dialog on and off).
→
Tap on
OK
to close the dialog field.
If an alarm is triggered in the data logger, an acoustic signal will
sound (provided that the signal transmitter is switched on) and the
ALARM LED
5
will light up. The ALARM LED
5
will light until the
alarm state is cancelled.
The alarm will be listed in the alarm list featuring the following
information: Alarm occurrence time, time of alarm end, and the
corresponding measuring channel.
1. In the home screen, tap on Display
to open the alarm list.
2. Go to the tab alarm list
.
➢
The alarm list will display all alarms saved so far with the affected
measuring channels and the type of alarm in the column alarm, the
alarm occurrence time in the column alarm occurrence, and the time
of the alarm end in the column alarm end.
Display in column Alarm
Message type of alarm
LIMIT-LOW
Lower limit value undershoot
LIMIT-HIGH
Upper limit value overshoot
RADIOFAIL
Interruption of radio connection
U-SENSOR
Sensor supply voltage dropping
below a sufficient value
BREAK
Sensor break