8 Settings
- 71
–
Selection
Transmission power
in dBm
Transmission power in
mW
Low
0 dBm
1 mW
Medium
10 dBm
10 mW
Maximum
17 dBm
100mW
➢
The wireless sensor will send at the selected transmission power.
8.2.4 How to set the Measuring Cycle
During the measuring cycle, the measurement data will be stored
with a timestamp in the wireless sensor. The data will be sent to the
data logger according to the output cycle. The fastest measuring
cycle that can be set is 10 seconds.
The measuring cycle can be configured to be shorter than the output
cycle. In that case several measurement data per measuring channel
will be sent to the data logger in one output cycle. The measuring
cycle must be at least one tenth of the output cycle.
Especially for measurements with fast measuring cycles make sure
that the operating mode
Measure
is set (see chapter 8.1.6 How to
set the Operating Mode).
!
The wireless sensor needs to be connected to the data logger.
!
No measurement may be started.
1. In the home screen, tap on Sensor overview
.
2. Tap on the arrow at the end of the line of the wireless sensor, whose
measuring cycle you want to change.
3. Tap on the field next to
Measuring cycle.
4. Enter the desired measuring cycle.
➢
The measurement data will be stored in the wireless sensor according
to the desired measuring cycle. The data will be sent to the data
logger according to the output cycle.