Menu Power Supply
reserve sufficient current for the measuring instrument and sensors and to
avoid adversely affecting measuring quality.
With our two mains adapters (accessories) the following recharge times are
needed:
Order-No.
Max. current
Device status Charge current Charge time
ZA2690-NA
200mA
switched off
400mA
approx. 4h
in operation
160mA
approx. 10h
ZA2690-NA2
800mA
switched off
1600mA
approx. 1h
in operation
400mA approx. 2.5h
For the duration of the charging phase the green LED on the adapter remains
lit continuously; as soon as the battery is fully charged, the adapter switches
over to trickle charge and the LED starts flashing.
If the accumulators are discerned (contact),
there appears in the display:
Accumulators:
Ø
The accumulator capacity must be programmed:
Capacity:
1600mAh
In charging mode the charge status is displayed:
Charge mode:
Charge
The charging current is set automatically:
Charge current: 1.60 A
As soon as the accumulator is fully charged,
charging status displays:
Charge mode: full
the charge current switches over to trickle charge:
Charge current: 0.01 A
If the battery temperature is either too high or too
low for charging or if the mains adapter is too
weak etc., the charging mode displays, e.g.:
Charge mode: T too high
or a numbered error mistake:
Charge mode: Error 1
ALMEMO
®
2690-8
69