9. Measuring using Measuring-Menus
9.4.5 Averaging over the Measuring Time
To determine the average value of all meas. values, that were acqired over the
conversion rate, over a defined period, the averaging mode
´CONT´
must be
set for the required measuring channel. Averaging can be carried out with or
without cycle. A measuring point scan will be performed in any case on start
and stop allowing to record the start and end values including the time of day.
A function channel
M(t)
is required for the recording of the average value M
(s. 10.3.9, 10.3.10).
Set the averaging mode:
Averaging Mode: CONT
Automatic clearing of average val. on start (s.
10.5.8
)
Check:
or, after selecting the average value, by using:
<CLR>
Starting the averaging using the key:
<START>
©
»
Reading the meas. time (s. 9.4.6) in function:
Meas.Time: 00:01:23.40
Stopping the averaging using the key:
<STOP>
ll
Reading the average value in function:
Average value: 13.24°C
Output of all function values of the menu by using:
<PRINT>
9.4.6 Measurement Time
For the averaging over time (see above) and for many other measurement ex-
periments, in many cases, the actual measurement time, from start to stop, is
required. The function ´Measuring time´ has a resolution of 0.10s and is availa-
ble to allow a continuous monitoring of the measurement time without clearing
the real time. If the function ´Clear Meas. Values On Start of a Measurement´
is activated within the operating parameters (s. 10.5.8), the measurement time
will also be automatically cleared on each start.
Function measuring time:
Measuring time: 00:00:00.00
Clearing the measuring time by using:
<CLR>
34 ALMEMO 2690-8
Conv. rate m
i
M
Start
Stop
M
=
∑
i
m
i
/
N