Measuring point scans and Output
The key
<FORM>
can be used as the quickest way to set the required Output
Format (printouts s. man. 6.6.1).
Change format:
<FORM>
Format columns next to each other ´
n
´:
Cycle Timer:
00:02:00
Sn
Change format:
<FORM>
Format table ´
t
´:
Cycle Timer:
00:02:00
St
Starting a cyclic measuring point scan:
<START>
For control purposes the following symbols will
continuously
, i.e. for the whole
meas.period, be indicated in the
status line
:
The start arrow lights up
´
©
´
Lightening up in case of a data output via interface
´
COM
´
Indicated when measured values are being stored (s. 10.1.2)
´
REC
´
Stopping a cyclic measuring point scan :
<STOP>
´
ll
´
9.3.3 Memory space, Memory Output, Clearing the Memory
The function
MemoryFreei
allows, during recording of measuring data, to con-
tinuously monitor the available memory space. By selecting this function, two
soft keys are available for a direct output and clearing of the memory. The out-
put format corresponds to the setting in the cycle (s. 9.3.2 and 10.1.2)
Function
MemoryFree
e.g.:
MemoryFree:
0378.4
kB
Memory data output (s. 10.2.5):
<PRINT>
Clear memory:
<CMEM>
9.3.4 Output of Menu Functions
Each data menu, together with all displayed functions, can be output via in-
terface to a printer or computer (connection of peripheral devices, see manual
5.2). If you have selected the standard display, pressing the key
<PRINT>
leads
to a printout of the following listing:
Print data menu:
<PRINT>
Meas. point, meas. value, designation:
01:+0023.5 °C Temperature
MAXIMUM: 01:+0020.0 °C
MINIMUM: 01:-0010.0 °C
PRINTTIMER: 00:01:23
Total memory capacity, free mem. in kB
MEMORY: S0512.1 F0324.4 A
The listing of the individual functions is given in section 6.6.1.
ALMEMO
®
2690-8
29