![Agilent Technologies OmniBER OTN J7230B Performance Verification Manual Download Page 93](http://html1.mh-extra.com/html/agilent-technologies/omniber-otn-j7230b/omniber-otn-j7230b_performance-verification-manual_2868192093.webp)
90
Verification Manual
1
Performance Tests
36
step 23
OC-12/STM-4 Mask
Pass/Fail
36
step 25
OC-12/STM-4 (ER)
>8.2 dB
N/A
38
step 33
OC-48/STM-16 Mask
Pass/Fail
39
step 35
OC-48/STM-16 (PWR)
–2.5 dBm
+4 dBm
39
step 36
OC-48/STM-16 (ER)
>8.2 dB
Pass/Fail
N/A
1550nm Optical Port (9.95 Gb/s and 10.7 Gb/s)
J7230B and J7232A Opt 108 only
41
step 8
OC-192/STM-64 Mask
Pass/Fail
41
step 10
OC-192/STM-64
(PWR)
–1 dBm
+2 dBm
42
step 11
OC-192/STM-64 (ER)
>8.2 dB
Pass/Fail
Option 112
43
step 17
OTU-2 Output Mask
Pass/Fail
43
step 19
OTU-2 Output (PWR)
–1 dBm
+2 dBm
43
step 19
OTU-2 Output (ER)
>8.2 dB
Pass/Fail
Optical Receiver Sensitivity, Optical Power and Frequency Measurement
Optical Receive Port (52 & 155 Mb/s)
All options
49
step 5
Laser LED ON
Pass/Fail
49
step 6
Note Attenuator
Setting
N/A
N/A
49
step 9
‘No Trouble’ Min Sens
Pass/Fail
50
step 10
Additional Attenuation
Typ > 1 dB
N/A
50
step 11
–30 dBm PWR meas
–31 dBm
–29 dBm
50
step 11
OC-1/STM-0
Frequency
51.83999 MHz
51.84001 MHz
51
step 12
–10 dBm PWR meas
–11 dBm
–9 dBm
Page No.
Test Description
Min
Result
Actual
Max
Verification.book Page 90 Monday, January 19, 2004 1:40 PM
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com