![Agilent Technologies OmniBER OTN J7230B Performance Verification Manual Download Page 33](http://html1.mh-extra.com/html/agilent-technologies/omniber-otn-j7230b/omniber-otn-j7230b_performance-verification-manual_2868192033.webp)
30
Verification Manual
1
Performance Tests
25
Set the PDH/DSn test set as follows:
TRANSMIT/RECEIVE settings to COUPLED
SIGNAL to DS1 rate
CLOCK to INTERNAL
OUTPUT LEVEL to DSX-1
LINE CODE to B8ZS
PAYLOAD TYPE to ESF UNSTRUCTURED
PATTERN to ALL ONES
26
Check there are no errors or alarms detected by the test set
receiver
27
Set the test set to frequency measurement results and check
that the displayed frequency is between:
1543999 Hz and 1544001Hz
28
Confirm that the OmniBER OTN is locked to the external
signal by disconnecting/re-connecting the BNC cable at the
OmniBER OTN CLOCK IN port.
29
Test is now complete.
PDH/DSn Test Set
OmniBER OTN
Bantam to Bantam cables
DS1 OUT
DS1 IN
DS1 OUT
DS1 IN
Verification.book Page 30 Monday, January 19, 2004 1:40 PM
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com