Performance Tests
8 Mb/s Output Mask Tests
1
Press
<Menu>
, select
Tx/Rx > Transmitter Settings
and set
the
Signal Mode
to
PDH/DSn
. Now set the
Signal Rate
to
8
Mb/s
.
2
Load the
8Mb_ITU.msk
test mask and check the Positive and
Negative pulses are within the mask (see examples).
.
Load the
8Mb_ITU.msk
test mask and check the Positive and Negative pulses are within the mask (see examples).
N O T E
To test Negative pulse set the Scope
Mask Test Set-up
panel to
Invert
Mask