![Agilent Technologies OmniBER OTN J7230B Performance Verification Manual Download Page 50](http://html1.mh-extra.com/html/agilent-technologies/omniber-otn-j7230b/omniber-otn-j7230b_performance-verification-manual_2868192050.webp)
Performance Tests
1
Verification Manual
47
Procedure
OC-1/STM-0 and OC-3/STM-1 Rates
1
Press
<Menu>
, choose
System > Stored Settings
then press
<Select>
. Select
Default
then
Recall
.
2
Press
<Menu>
, choose
Tx/Rx > Coupling
then press
<Select>
. Select
Copy Tx to Rx
.
3
Connect the test set-up as shown below depending on the
wavelengths installed.
WA R N I N G
Safety precautions must be observed when handling the OmniBER
OTN’s Optical Modules as these generate laser signals which can
cause serious injury. The guidelines below must be followed:
•
Check the connector configuration of the Fiber Optic Interfaces. If non FC/PC
connectors are fitted then remove them, then fit the FC/PC connector interface.
•
Check for any damage to the OmniBER OTN’s Fiber Optic Interface spring
loaded aperture covers and connectors. Do not power up the instrument if in
any doubt about the integrity of these connectors.
Verification.book Page 47 Monday, January 19, 2004 1:40 PM
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com