Press
<Smart Test>
, choose
Shortcuts > Frequency
Measurement
then press
<Select>
. The OmniBER OTN now
displays the received frequency results.
Check that the frequency measurement reads between:
51.83999 and 51.84001 MHz for [STM-0 OPT] or [OC-1]
(155.51999 and 155.52001 MHz for [STM-1 OPT] or [OC-3]
)
12
Reduce the optical attenuator setting by a further 20 dB and
check the receiver
Optical Power
result is between:
–11 dBm and –9 dBm
Verification.book Page 51 Monday, January 19, 2004 1:40 PM