Performance Test Record
Model: Agilent J7231B/J7232A OmniBER OTN
Location:
Serial No.:
Tested by:
Options:
Temperature:
Certified by:
Humidity:
Date:
N O T E
The test steps applicable are dependent on the instrument options
installed. Enter N/A if a test section does not apply for your
instrument.
Page No.
Test Description
Min
Result
Actual
Max
Performance Self Test
15
step 5
Confidence Test
Pass/Fail
15
step 6
Frequency
Measurement
Pass/Fail
Internal Reference Clock Accuracy
20
step 4
2 MHz Clock OUT
2.0479908 MHz
2.0480092 MHz
22
step 13
DS1 Clock OUT
–4.5 ppm
+4.5 ppm
23
step 2
Alternative DS1 Test
(Counter indication for
framed signal)
1.540993 MHz
1.541007 MHz
External Clock Reference Inputs and Clock Reference Output
26
step 6
2 MHz Clock Ext
2.047999999 MHz
2.048000001 MHz
27
step 7
Signal Locked
Pass/Fail
27
step 10
/
step
6
10 MHz Clock Ext
2.047999999 MHz
2.048000001 MHz
Verification.book Page 88 Monday, January 19, 2004 1:40 PM