6 Language Dictionary
76
Series N6700 User’s Guide
SCPI Command Summary
Subsystem Commands
NOTE
Some [optional] commands have been included for clarity. All settings commands
have a corresponding query. Not all commands apply to all models.
SCPI Command
Description
ABORt
:ACQuire (@chanlist)
Resets the measurement trigger system to the Idle state
:TRANsient (@chanlist)
Resets the transient trigger system to the Idle state
CALibrate
:CURRent
[:LEVel] <NRf>, (@channel)
Calibrates the output current programming
:MEASure <NRf>, (@channel)
Calibrates the current measurement
:PEAK (@channel)
Calibrates the peak current limit (Agilent N6751A/52A/61A/62A)
:DATA <NRf>
Enters the calibration value
:DATE <SPD>, (@channel)
Sets the calibration date
:DPRog (@channel)
Calibrates the current downprogrammer
:LEVel P1 | P2 | P3
Advances to the next calibration step
:PASSword <NRf>
Sets the numeric calibration password
:SAVE
Saves the new cal constants in non-volatile memory
:STATE <Bool> [,<NRf>]
Enables/disables calibration mode
:VOLTage
[:LEVel] <NRf>, (@channel)
Calibrates the output voltage programming
:CMRR (@channel)
Calibrates common mode rejection ratio (N6751A/52A/61A/62A)
:MEASure <NRf>, (@channel)
Calibrates the voltage measurement
DISPlay[:WINDow]:VIEW METER1 | METER4
Selects 1-channel or 4-channel meter view
FETCh (Note 1) | MEASure
[:SCALar]
:CURRent [:DC]? (@chanlist)
Returns the average output current
:VOLTage [:DC]? (@chanlist)
Returns the average output voltage
:ARRay
(Array commands only on Agilent N6761A/62A and Option 054)
:CURRent [:DC]? (@chanlist)
Returns the instantaneous output current
:VOLTage [:DC]? (@chanlist)
Returns the instantaneous output voltage
INITiate
[:IMMediate]
(Acquire command only on Agilent N6761A/62A and Option 054)
:ACQuire (@chanlist)
Enables the measurement system to receive triggers
:TRANsient (@chanlist)
Enables the output transient system to receive triggers
:CONTinuous
:TRANsient <Bool>, (@chanlist)
Enables/disables continuous transient triggers
OUTPut
[:STATe] <Bool> [,NORelay], (@chanlist)
Enables/disables the specified output channel(s)
:DELay
:FALL <NRf+>, (@chanlist)
Sets the output turn-off sequence delay
:RISE <NRf+>, (@chanlist)
Sets the output turn-on sequence delay
:INHibit
:MODE LATChing | LIVE | OFF
Sets the remote inhibit input
Summary of Contents for N6700B
Page 1: ...A A A A Agilent Technologies Modular Power System Series N6700 User s Guide...
Page 28: ......
Page 160: ......