Verification and Calibration Appendix B
Series N6700 User’s Guide
157
Step 29.
Disconnect the Agilent 3458A from the output terminals. Then select
the second current calibration point. Wait a minimum of 5 minutes
for the internal temperature to stabilize.
Front Panel:
SCPI Command:
Select Next. The information field
should indicate: Enter P2 data
CAL:LEV P2
Step 30.
Then select the third current calibration point.
Front Panel:
SCPI Command:
Select Next. The information field
should indicate: Enter P3 data
CAL:LEV P3
*OPC?
Low Range Current Programming Calibration
This only applies to Agilent Models N6761A and N6762A.
Step 31.
Connect only the current measurement terminals of the Agilent
3458A to the output terminals.
Step 32.
Select the low-current programming range. The following example
selects the 0.1 amp low-current programming range of channel 1. The
value to program a range must be the maximum current of the range.
Front Panel:
SCPI Command:
Select
System\Admin\Cal\Function\Iprog
.
Then select the Low range.
CAL:CURR 0.1
, (@1)
Step 33.
Select the first current calibration point. Wait a minimum of 5
minutes for the output current to stabilize.
Front Panel:
SCPI Command:
Select Next. The information field
should indicate: Enter P1 data
CAL:LEV P1
*OPC?
Step 34.
Measure the current with the
Agilent 3458A and enter the value.
Front Panel:
SCPI Command:
Select the Measured Data field. Enter
the data from the external DVM.
Press Select when done.
CAL:DATA <data>
Step 35.
Select the second current calibration point. Then wait 30 seconds
for the output current to stabilize.
Front Panel:
SCPI Command:
Select Next. The information field
should indicate: Enter P2 data
CAL:LEV P2
*OPC?