Appendix D Service
190
Series N6700 User’s Guide
Query
Errors
(these errors set Standard Event Status register bit #2)
−
400
Query Error
Generic error query
−
410
Query INTERRUPTED
A condition causing an interrupted query error occurred.
−
420
Query UNTERMINATED
A condition causing an unterminated query error occurred.
−
430
Query DEADLOCKED
A condition causing a deadlocked query error occurred.
−
440
Query UNTERMINATED after indefinite response
A query was received in the same program message after a query indicating an
indefinite response was executed.
Selftest
Errors
(these errors set Standard Event Status register bit #3)
202
Selftest Fail Aux Adc 0 expected <n1> to <n2>, measured <n3>, chan <n4>
Auxiliary ADC failed. n1 and n2 are the expected limits. n3 is the measured value. n4 is the
channel location of the failed module.
202
Selftest Fail DACs 0 expected <n1> to <n2>, measured <n3>, chan <n4>
Both voltage and current DACs are at zero. n1 and n2 are the expected limits. n3 is the
measured value. n4 is the channel location of the failed module.
202
Selftest Fail DACs 1 expected <n1> to <n2>, measured <n3>, chan <n4>
Voltage DAC is at zero; current DAC is at full scale. n1 and n2 are the expected limits. n3 is the
measured value. n4 is the channel location of the failed module.
202
Selftest Fail DACs 2 expected <n1> to <n2>, measured <n3>, chan <n4>
Voltage DAC is at full scale; current DAC is at zero. n1 and n2 are the expected limits. n3 is the
measured value. n4 is the channel location of the failed module.
202
Selftest Fail DACs 3 expected <n1> to <n2>, measured <n3>, chan <n4>
Both voltage and current DACs are at full scale. n1 and n2 are the expected limits. n3 is the
measured value. n4 is the channel location of the failed module.