Appendix A Specifications
122
Series N6700 User’s Guide
Agilent Models N6751A/N6752A and N6761A/N6762A (continued)
N6751A / N6752A
N6761A / N6762A
Maximum Up-programming Time with full resistive load:
(time f
rom 10% to 90% of total voltage excursion)
Voltage setting from 0 V to 10 V
0.2 ms
0.6 ms
Voltage setting from 0 V to 50 V
1.5 ms
2.2 ms
Maximum Up-programming Settling Time with full resistive load:
(time from start of voltage change to within 50mV of final value)
Voltage setting from 0 V to 10 V
0.5 ms
0.9 ms
Voltage setting from 0 V to 50 V
4.0 ms
4.0 ms
Maximum Down-programming Time with no load:
(time f
rom start of voltage change to output voltage < 0.5 V)
Voltage setting from 10 V to 0 V
0.3 ms
0.3 ms
Voltage setting from 50 V to 0 V
1.3 ms
1.3 ms
Maximum Down-programming Settling Time with no load:
(time f
rom start of voltage change to output voltage within 50mV of final value)
Voltage setting from 10 V to 0 V
0.45 ms
0.45 ms
Voltage setting from 50 V to 0 V
1.4 ms
1.4 ms
Down-programming Time with 1000 µF load:
NOTE 2
(time f
rom start of voltage change to output voltage < 0.5 V)
Voltage setting from 10 V to 0 V
2.1 ms
4.5 ms
Voltage setting from 50 V to 0 V
11 ms
23 ms
Down-programming Capability:
Continuous power
7 W
7 W
Peak current
7 A
3.8 A
Remote Sense Capability:
Outputs can maintain specifications with up to a 1-volt
drop per load lead.
Series and Parallel Operation:
Identically rated outputs can be operated directly in
parallel or can be connected for straight series
operation. Auto-series and auto-parallel operation is not
available.
Load Cross Regulation:
Voltage, from no load to full load
1 mV
0.5 mV
Current, from no load to full load
1 mA
5
µ
A
2
Modules can discharge a 1000
µ
F capacitor from 50V to 0V at a rate of 4 times/second.