Replacing Assemblies
5
7000 Series Oscilloscopes Service Guide
93
To remove the Front Panel assembly
1
Remove cable shield from front deck. Carefully squeeze
the shield so that it clears the sheet metal tabs holding it
place then slide it free of the front deck:
Figure 25
Removing cable shield
2
Remove keyboard cables that extend thru front deck.
Figure 26
Removing keyboard cables
WA R N I N G
Thin sheet metal parts may have sharp edges. Handle with care to avoid injury.