7000 Series Oscilloscopes Service Guide
2
Testing Performance
Figure 3
Setting up Equipment for Voltage Measurement Accuracy
Test
4
Adjust the output so that the multimeter reading displays
the first Volts/div calibrator setting value in
Table 7
.
Wait a few seconds for the measurement to settle.
5
Press the
Y2
softkey, then position the Y2 cursor to the
center of the voltage trace using the Entry knob.
The
Δ
Y value on the lower line of the display should be
within the test limits of
Table 7
. If a result is not
within the test limits, go to the “Troubleshooting”
chapter. Then return here.
6
Continue to check the voltage measurement accuracy with
the remaining Volts/div setting values in
Table 7
.
Oscilloscope
Oscilloscope
Calibrator
Digital
Multimeter
BNC Tee
BNC (f) to dual
banana adapter