16
7000 Series Oscilloscopes Service Guide
1
Characteristics and Specifications
Transient Withstand Capability
C A U T I O N
Maximum input voltage for analog inputs
CAT I 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk
with 10073C or 10074C 10:1 probe: CAT I 500 Vpk
C A U T I O N
Do not exceed 5 Vrms in 50
Ω
mode. Input protection is enabled in
50
Ω
mode, and the 50
Ω
load will disconnect if greater than 5 Vrms is
detected. However, the input could still be damaged, depending on the
time constant of the signal. The 50
Ω
input protection mode only
functions when the oscilloscope is powered on.
C A U T I O N
Maximum input voltage for logic channels:
±40 V peak CAT I; transient overvoltage 800 Vpk