Testing Performance
2
7000 Series Oscilloscopes Service Guide
55
1
Connect the equipment (see
Figure 8
).
a
Use the N cable to connect the signal generator to the
power splitter input.
b
Connect one output of the power splitter to the Aux
Trig input through a 50
Ω
feedthrough termination.
c
Connect the power sensor to the other output of the
power splitter.
Figure 8
Setting Up Equipment for 4-Channel External Trigger Sensitiv-
ity Test
Oscilloscope
Signal
Generator
Power Meter
Power Splitter
Power Sensor
N to BNC adapter
N Cable
50
Ω
Feedthrough