34
7000 Series Oscilloscopes Service Guide
2
Testing Performance
1
Set up the oscilloscope.
a
Adjust the channel 1 position knob to place the
baseline at 0.5 major division from the bottom of the
display.
b
Set the Volts/Div setting to the value in the first line in
Table 7
.
Table 6
Equipment Required to Verify Voltage Measurement Accuracy
Equipment
Critical Specifications
Recommended
Model/Part
Oscilloscope
Calibrator
14 mV to 35 Vdc,
0.1 V resolution
Fluke 5820A
Digital multimeter
Better than 0.01% accuracy
Agilent 34401A
Cable
BNC, Qty 2
Agilent 10503A
Shorting cap
BNC
Agilent 1250-0774
Adapter
BNC (f) to banana (m)
Agilent 1251-2277
Adapter
BNC tee (m) (f) (f)
Agilent 1250-0781 or
Pomona 3285
Blocking capacitor
Agilent 10240-70001 +
Agilent 1250-0774 or
Agilent Pomona
4288 + Pomona 5088A