58
7000 Series Oscilloscopes Service Guide
2
Testing Performance
Horizontal
Δ
t Accuracy
Generator Setting
Test Limits
Results
Period
100
μ
s
99.8
μ
s to 100.2
μ
s
________
Period
100 ns
99.8 ns to 100.2 ns
________
Period
5 ns
4.96 ns to 5.04 ns
________
Trigger Sensitivity
Test Limits
Channel 1
Channel 2 Channel 3* Channel 4*
Internal trigger
1 GHz models
1 division at 1 GHz
________ ________ ________ _______
500 MHz models
1 division at 500 MHz
________ ________ ________ _______
350 MHz models
1 division at 350 MHz
________ ________ ________ _______
100 MHz models
1 division at 100 MHz
________ ________ ________ _______
Trigger Sensitivity
Test Limits
Channel 1
Channel 2 Channel 3* Channel 4*
External trigger (7052, 7032, 7012)
± 1 V range:
100 mV at 25 MHz
________ ________ ________ ________
7052
200 mV at 500 MHz
________ ________ ________ ________
7032
200 mV at 350 MHz
________ ________ ________ ________
7012
200 mV at 100 MHz
________ ________ ________ ________
External trigger (7052, 7032, 7012)
± 8 V range:
250 mV at 25 MHz
________ ________ ________ ________
7052
500 mV at 500 MHz
________ ________ ________ ________
7032
500 mV at 350 MHz
________ ________ ________ ________
7012
500 mV at 100 MHz
________ ________ ________ ________
External trigger (7104, 7054, 7034, 7014)
500 mV at 500 MHz
________ ________ ________ ________
* Where applicable