Calibrating and Adjusting
3
7000 Series Oscilloscopes Service Guide
63
BNC(f)- to- BNC(f) (barrel connector) to the tee as
shown below.
Figure 10
User Calibration cable for 4-channel oscilloscope
3
Connect a BNC cable (40 inches maximum) from the TRIG
OUT connector on the rear panel to the BNC barrel
connector.
4
Press the
Utility
key, then press the
Service
softkey.
5
Begin the Self Cal by pressing the
Start User Cal
softkey.
6
When the User Cal is completed, set the rear- panel
CALIBRATION switch to PROTECTED.
Longer cable
to TRIG OUT
To Channel 3
To Channel 2
To Channel 1
To Channel 4