52
7000 Series Oscilloscopes Service Guide
2
Testing Performance
1
Connect the equipment (see
Figure 7
).
a
Use the N cable to connect the signal generator to the
power splitter input.
b
Connect one output of the power splitter to the Ext
Trigger input.
c
Connect the power sensor to the other output of the
power splitter.
Figure 7
Setting Up Equipment for 2-Channel External Trigger Sensitiv-
ity Test
Oscilloscope
Signal
Generator
Power Meter
Power Splitter
Power Sensor
N to BNC adapter
N Cable
50
Ω
Feedthrough