86
1260 SFC Binary Pump
6
Error Information
Module Error Messages
Pump Head Missing
Error ID: 2202, 2212
Pump Head Missing:
Pump channel A
B: Pump Head Missing:
Pump channel B
The pump-head end stop in the pump was not found.
When the pump restarts, the metering drive moves forward to the mechanical
end stop. Normally, the end stop is reached within 20 s, indicated by an
increase in motor current. If the end point is not found within 20 s, the error
message is generated.
Index Limit
Error ID: 2203, 2213
Index Limit:
Pump channel A
B: Index Limit:
Pump channel B
The time required by the piston to reach the encoder index position was too
short (pump).
During initialization, the first piston is moved to the mechanical stop. After
reaching the mechanical stop, the piston reverses direction until the encoder
index position is reached. If the index position is reached too fast, the error
message is generated.
Probable cause
Suggested actions
1
Pump head not installed correctly (screws
not secured, or pump head not seated
correctly).
Install the pump head correctly. Ensure nothing
(e.g. capillary) is trapped between the pump
head and body.
2
Broken piston.
Exchange the piston.
Probable cause
Suggested actions
1
Irregular or sticking drive movement.
Remove the pump head, and examine the
seals, pistons, and internal components for
signs of wear, contamination or damage.
Exchange components as required.
2
Defective pump drive assembly.
Please contact your Agilent service
representative.
Summary of Contents for G4782A
Page 8: ...8 1260 SFC Binary Pump Contents...
Page 22: ...22 1260 SFC Binary Pump 1 Introduction Leak and Waste Handling...
Page 30: ...30 1260 SFC Binary Pump 2 Site Requirements and Specifications Performance Specifications...
Page 66: ...66 1260 SFC Binary Pump 5 Troubleshooting and Diagnostics Agilent Lab Advisor Software...
Page 140: ...140 1260 SFC Binary Pump 9 Parts and Materials for Maintenance Accessory Kit...
Page 172: ...172 1260 SFC Binary Pump 11 Hardware Information Instrument Layout...