146
1260 SFC Binary Pump
10
Identifying Cables
Remote Cables
Remote Cables
ERI (Enhanced Remote Interface)
5188-8029 ERI to general purpose
5188-8044 ERI to ERI (Connector D_Subminiature 15 pin)
p/n 5188-8029
pin
Color code
Enhanced
Remote
Classic
Remote
Active (TTL)
1
white
IO1
START
REQUEST
Low
2
brown
IO2
STOP
Low
3
green
IO3
READY
High
4
yellow
IO4
POWER ON
High
5
grey
IO5
NOT USED
6
pink
IO6
SHUT DOWN
Low
7
blue
IO7
START
Low
8
red
IO8
PREPARE
Low
9
black
1wire DATA
10
violet
DGND
11
grey-pink
+5V ERI out
12
red-blue
PGND
13
white-green
PGND
14
brown-green
+24V ERI out
15
white-yellow
+24V ERI out
NC
yellow-brown