1260 SFC Binary Pump
105
Maintenance
8
Remove the Pump Head Assembly
Remove the Pump Head Assembly
When
•
Exchanging pump seals
•
Exchanging pistons
•
Exchanging seals of the seal wash option
Tools required
p/n
Description
8710-0510
Wrench open 1/4 — 5/16 inch
8710-2411
Hex key 3 mm12 cm long
8710-2392
Hex key 4 mm15 cm long T-handle
5023-0240
Hex driver, ¼", slitted
Preparations
Switch off the pump at the main power switch.
C A U T I O N
Damage of the pump drive
Starting the pump when the pump head is removed may damage the pump drive.
➔
Never start the pump when the pump head is removed.
N O T E
Both pump head assemblies use the same internal components. In addition, pump head A
is fitted with the purge valve. The following procedure describes the removal and
disassembly of pump head A (left). For pump head B (right) proceed in the same way and
skip steps that deal with the purge valve.