1260 SFC Binary Pump
71
Error Information
6
General Error Messages
Remote Timeout
Error ID: 0070
A not-ready condition is still present on the remote input. When an analysis is
started, the system expects all not-ready conditions (for example, a not-ready
condition during detector balance) to switch to run conditions within one
minute of starting the analysis. If a not-ready condition is still present on the
remote line after one minute the error message is generated.
Lost CAN Partner
Error ID: 0071
During an analysis, the internal synchronization or communication between
one or more of the modules in the system has failed.
The system processors continually monitor the system configuration. If one or
more of the modules is no longer recognized as being connected to the system,
the error message is generated.
Probable cause
Suggested actions
1
Not-ready condition in one of the
instruments connected to the remote line.
Ensure the instrument showing the not-ready
condition is installed correctly, and is set up
correctly for analysis.
2
Defective remote cable.
Exchange the remote cable.
3
Defective components in the instrument
showing the not-ready condition.
Check the instrument for defects (refer to the
instrument’s documentation).
Probable cause
Suggested actions
1
CAN cable disconnected.
•
Ensure all the CAN cables are connected
correctly.
•
Ensure all CAN cables are installed
correctly.
2
Defective CAN cable.
Exchange the CAN cable.
3
Defective main board in another module.
Switch off the system. Restart the system, and
determine which module or modules are not
recognized by the system.
Summary of Contents for G4782A
Page 8: ...8 1260 SFC Binary Pump Contents...
Page 22: ...22 1260 SFC Binary Pump 1 Introduction Leak and Waste Handling...
Page 30: ...30 1260 SFC Binary Pump 2 Site Requirements and Specifications Performance Specifications...
Page 66: ...66 1260 SFC Binary Pump 5 Troubleshooting and Diagnostics Agilent Lab Advisor Software...
Page 140: ...140 1260 SFC Binary Pump 9 Parts and Materials for Maintenance Accessory Kit...
Page 172: ...172 1260 SFC Binary Pump 11 Hardware Information Instrument Layout...