110
1260 SFC Binary Pump
8
Maintenance
Maintenance of a Pump Head with Seal Wash
Maintenance of a Pump Head with Seal Wash
When
In case of maintenance or pump head internal leaks
Tools required
p/n
Description
8710-2411
Hex key 3 mm12 cm long
8710-2392
Hex key 4 mm15 cm long T-handle
01018-23702 Insert
tool
Screwdriver, small flat head
Parts required
p/n
Description
0905-1718
Wash seal (PTFE)
5062-2484
Gasket, seal wash (pack of 6)
5067-4695
Sapphire piston
0905-1719
PE Seal
Preparations
•
Switch off the pump at the main power switch.
•
Open the doors of the module.
•
Use an optional solvent shutoff valve or lift up solvent filters for avoiding leakages.
•
Remove the Pump Head Assembly.
•
Remove the wash solvent tubings from the support ring inlet and outlet.
1
Place the pump head on a flat surface. Loosen the lock
screw (two turns).
2
While holding the lower half of the assembly (piston
housing), carefully pull the pump housing away from the
piston housing.