You can also set Velocity Factor by pressing
Meas Setup 4
then
Calibration
(settings).
Increase Dynamic Range
Dynamic range is the difference between maximum input power to the FieldFox
receiver (without compressing the receiver), and the minimum measurable
power (noise floor). Measurement accuracy is increased when the DUT response
is at least 10 dB above the noise floor. For a measurement to be valid, input
signals must be within these boundaries.
The following settings will increase the dynamic range of your
NA mode
measurement.
Increase Power Level: Press
Meas 4
then
Output Power High
Lower the IFBW: Press
BW 2
then
IF BW
Increase Averaging: Press
BW 2
then
Average
The following procedure MAY increase the dynamic range of your NA mode
measurement. The results you see will depend on the performance of your DUT.
With an S21 trace active:
1. With RF OUT (port-2) open, press
Trace 6
then
Math and Memory
then
Data->Mem
2. Re-connect the DUT.
3. Press
Data Math
then
Data-Mem