System Settings
205
Startup Mode
The Startup Mode setting determines the mode that becomes active (visible)
when User Preset is set to OFF (Factory Preset) and the FieldFox is powered ON.
Learn more about User Preset on page 200.
How to set Startup Mode
Press
System 7
Then
Preferences
Then
Preferences
Then scroll to
Startup Mode
and press
Edit
.
Then repeatedly press the
Startup Mode
to scroll through the available
choices.
o
<Default>
is the factory default mode for your FieldFox model.
o
Built-in Power Meter Mode is NOT allowed as a Startup Mode.
Battery Saver
When the battery saver is OFF, the RF source power is left ON at the end of the
sweep which consumes more battery power. This feature is used in CAT, NA,
VVM modes, and SA mode when the internal source is turned ON.
Battery Saver is the ONLY FieldFox setting that must be made as a Preference
from the Quick Settings table. This setting remains until you change it again.
How to set Battery Saver
Press
System 7
Then
Preferences
Then
Preferences
Then scroll to
Battery Saver
o
ON
– Default setting. To conserve battery power, the RF source power is
turned OFF at the end of each CAT, NA, and VVM mode sweep. This
behavior does not impact FieldFox measurements.
o
OFF
– Source power is kept ON at all times, including during trigger Hold
mode and between single sweep acquisitions. The RF source stays at the stop
frequency until another sweep begins. This provides a constant,
uninterrupted RF output from the RF OUT port. This behavior also provides
the highest trace stability sweep-to-sweep which may only be evident while
using either single sweep or Run/Hold. With Battery Saver OFF, the use of a
full battery charge is reduced by approximately 45 minutes.