Installation Note E8364-90022
27
Install the New A27, A28, A29, and A30 First Converters (Mixers)
1. Remove all four mixers from the previously removed mixer bracket by removing three
screws (item
①
) from each mixer. Retain the screws.
2. Remove the mixer bias board by removing the four attachment screws (item
②
). Retain the
screws.
3. Install the new mixer bias board onto the bracket by
loosely
installing the attachment
screws (item
②
) from the old mixer bias board. Do not tighten the screws at this time.
4. Using a T-10 TORX driver, install the four new mixers onto the bracket using the screws
(item
①
) from the previously removed mixers. Be sure to orient the mixers as shown, and
be sure that the DC feed-through pins align properly with the mixer bias board.
5. Using a T-10 TORX driver, tighten the mixer bias board attachment screws (item
②
).
6. The mixer bracket with new mixers attached is now ready to be installed in the analyzer.
Figure 19
Mixers Bracket Assembly