16
Installation Note E8364-90022
Remove the A18 MA 20
1. Using a 5/16-inch wrench, disconnect the two semirigid cables (item
①
) from the A18 MA
20. Remove these cables completely to allow the A18 MA 20 to be removed from the
analyzer. Retain these cables for reinstallation later.
2. Disconnect the ribbon cable (item
②
) at the A18 MA 20.
3. Using a T-10 TORX driver, remove the four mounting screws (item
③
) from the side of the
analyzer. Retain the mounting screws for reinstallation later.
CAUTION
Be careful not to damage the center pins of the semirigid cables. Some flexing of
the cables is necessary to remove the assembly. Do not over-bend them.
4. Remove the A18 MA 20 from the analyzer. Observe the
CAUTION
above.
Figure 8
A18 MA 20 Removal