Installation Note E8364-90022
23
Install the New A25 and A26 Test Port Couplers
1. Position the A25 and A26 test port couplers in the analyzer as shown.
2. Install the test port coupler nuts (item
①
) and, with a 1-inch torque wrench, tighten them
to 72 in-lbs.
3. The cables will be installed later.
Figure 15
A25 and A26 Test Port Couplers Installation