Installation Note E8364-90022
11
Step 3. Replace the A12 Source 20
1. With a 5/16-inch wrench, disconnect semirigid cable (item
①
).
2. Disengage the A12 source 20 from the A11 connector and slide it out of the analyzer while
observing the
CAUTION
below.
CAUTION
To avoid damage to the semirigid cable (item
①
), pull down on the cable from the
bottom side of the analyzer to move it away while sliding out the A12 source 20.
The cable has some spring to it and the RF pin could be damaged by the A12
source 20.
Figure 4
A12 Source 20 Replacement
NOTE
The figure above shows the E8364B front panel and floppy disk drive. The
E8364C front panel has a slightly different appearance and does not
include a floppy disk drive.