Installation Note E8364-90022
15
Step 6. Remove the Old Bottom Side Hardware and Cables
•
•
•
Remove the A22 Switch/Splitter
•
Remove the A23 and A24 Detectors
•
Remove the A27, A28, A29, and A30 First Converters (Mixers)
•
Remove the A25 and A26 Test Port Couplers
Remove the A17 LOMA 10
1. Using a 5/16-inch wrench, disconnect the three semirigid cables (item
①
) from the A17
LOMA 10. Two of these cables should already be disconnected (from step 4).
2. Disconnect the ribbon cable (item
②
) at the A17 LOMA 10.
3. Using a T-10 TORX driver, remove the four mounting screws (item
③
) from the side of the
analyzer. Retain the mounting screws for reinstallation later.
CAUTION
Be careful not to damage the center pins of the semirigid cables. Some flexing of
the cables is necessary to remove the assembly. Do not over-bend them.
4. Remove the A17 LOMA 10 from the analyzer. Observe the
CAUTION
above.
Figure 7
A17 LOMA 10 Removal