20
Installation Note E8364-90022
Remove the A25 and A26 Test Port Couplers
1. Using a 5/16-inch wrench, remove the two semirigid cables (item
①
) from each of the test
port couplers. Discard these cables, they will be replaced with new ones.
2. Using a 1-inch wrench, remove the test port coupler nuts (item
②
) from the front panel.
Retain the test port coupler nuts for reinstallation later.
CAUTION
Be careful not to damage the center pins of the semirigid cables. Some flexing of
the cables is necessary to remove the assembly. Do not over-bend them.
3. Remove the A25 and A26 test port couplers from the analyzer. Observe the
CAUTION
above.
Figure 12 A25 and A26 Test Port Couplers Removal