Installation Note E8364-90022
21
Step 7. Install the New Hardware and Cables
•
•
•
Install the New A25 and A26 Test Port Couplers
•
Install the New A23 and A24 SOMA 70s
•
•
Install the New A27, A28, A29, and A30 First Converters (Mixers)
•
Install the New A17 LOMA 12
1. Position the new A17 LOMA 12 in the analyzer as shown. Observe the
CAUTION
below.
CAUTION
Be careful not to damage the center pins of the semirigid cables. Some flexing of
the cables is necessary when installing the assembly. Do not over-bend them.
2. Using a T-10 TORX driver, install three of the four mounting screws (item
③
), previously
removed from the old A17 LOMA 10, through the side of the analyzer as shown. One of the
screws removed previously is not reused.
3. Using a 5/16-inch torque wrench (set to 10 in-lbs), reconnect the two semirigid cables
(item
①
) to the A17 LOMA 12. Observe the
CAUTION
above.
4. Reconnect the ribbon cable (item
②
) to the A17 LOMA 12.
Figure 13
A17 LOMA 12 Installation