Chapter 6: Maintenance and troubleshooting
BioCel User Guide
2.
In the
Robot Error
dialog box, select
Retry
.
Related topics
Resolving plate placing errors
About plate placing
errors
If the robot does not place a plate accurately on a device, the problem
could be with the:
❑
Teachpoint for either that device or for the previously scheduled
device, including approach height.
In this case, see for more information about fine-tuning teachpoints.
❑
Device, because it was moved or reconfigured by someone else who
did not re-teach the teachpoints.
❑
Plate.
In this case, replace the plate with one that is not deformed.
Relate information
For information about...
See...
Opening a door during a run
“Pausing and stopping a run” on
page 60
Safety interlocks and emergency
stops
“Comparing interlock and emergency
circuits” on page 62
For information about...
See...
Fine-tuning teachpoints