65
Chapter 4: Performing a run
BioCel User Guide
Related topics
Cleaning up after a run
About this topic
This topic provides a list of tasks to perform on the BioCel after the last
run. For more information about cleaning up after the run, see
VWorks
User Guide
.
Procedure
To clean up after a run:
1.
Remove sample plates that have been used.
2.
Remove manually placed plates from platepads, hotel shelves,
VPrep module shelves, and so on.
3.
If necessary, remove the counterbalance plate from VSpin devices.
4.
Wash all items, such as, reservoirs, tubes and manifolds, that were
contacted by liquid.
5.
Empty the waste bin.
6.
Check run logs.
Red
Error
During operation, an error
occurred and the system cannot
continue without user
intervention.
Yellow
Error
During operation, a network or
disk space error occurred. The
system will stop until the error is
fixed.
Light Pattern
Status
Comment
For information about...
See...
Setting error options
VWorks User Guide
Monitoring the UPS
VWorks User Guide
Troubleshooting an error
❑
“Maintenance and
troubleshooting” on page 85
❑
VWorks User Guide
Stopping a run
“Pausing and stopping a run” on
page 60
Setting up the waste bin
“Preparing devices and accessories” on
page 46
Using the User Message task
VWorks User Guide
Summary of Contents for BioCel 1200
Page 1: ...Agilent Technologies BioCel System User Guide...
Page 4: ...Letter to our Customers 2 Agilent Technologies Inc Automation Solutions Division...
Page 18: ...Chapter 1 Introduction BioCel User Guide 12...
Page 64: ...Chapter 3 Preparing for a run BioCel User Guide 58...
Page 74: ...Chapter 4 Performing a run BioCel User Guide 68...
Page 100: ...Chapter 6 Maintenance and troubleshooting BioCel User Guide 94...
Page 104: ...Index BioCel User Guide 98...
Page 105: ......